ATI Q46H-65 Chlorine Dioxide Monitor

Construction:



On-line Monitoring & Control for ClO2 Disinfection Systems!

Chlorine Dioxide is a powerful oxidizing agent used for disinfection of potable water, cooling towers, and food wash water. It is also widely used for the bleaching of pulp in paper mills. As a disinfectant, ClO2 has advantages over chlorine in that it does not react with ammonia and many organic compounds while proving more effective at controlling bacteria, spores, viruses, and cysts. It is particularly well suited for removing biofilms from the inside of piping systems.

 

ATI's Model Q46H/65 Chlorine Dioxide Monitor provides an economical and reliable measurement system for water quality monitoring and controlling ClO2 treatment systems. With a variety of outputs including 4-20 mA analog, PID control and digital communications, the Q46H/65 is adaptable to any ClO2 application.

 



 

INTERFERENCE-FREE
 
 

CIO2 MEASUREMENT

 

Q46H/65 monitors use a   polarographic membraned sensor to accurately measure ClO2 in water. The   sensor operates much like a battery, generating current that is linearly   proportional to the concentration of chlorine dioxide in solution. A   ClO2 permeable membrane isolates the sensor from the measured sample and   insures that the measurement is interference free.
 

 
Two   versions of the sensor are available, a sensor intended for installation in a   flowcell and a sensor intended for submersion applications. Requiring   minimal maintenance, sensors are easy to use and easy to maintain. An   integral RTD provides temperature compensation and allows temperature to   be displayed and transmitted from the monitor.


 


    Submersion & Flowcell
 
 

 

 

 

SENSOR STABILIZATION
 
Chlorine   Dioxide sensors require 2 to 4 hours of stabilization time when first installed   or after membrane change. ATI offers a battery powered “polarizer” that can   be used to stabilize a spare sensor so it is ready to run within a few   minutes of installation. Polarizers simply plug into the sensor connector and   require no adjustments.


 

 

ClO2 Monitoring with Direct Membrane Sensing!

 

SENSOR & FLOWCELL OPTIONS

Chlorine Dioxide sensors require a steady flow of sample across the membrane at the tip of the sensing assembly. ATI offers a number of options for flowcells, including the standard constant-head overflow system, a sealed flowcell for pressures up to 50 PSI, and a low-volume flowcell where low sample flow is important. A 1-½” flow tee is also available for in-line applications with reliable constant flow and pressure conditions. For simplicity of installation, complete flow control assemblies are available.

 

 

 

 

PRODUCT FEATURES

pH Input Option

Enables monitor to measure both chlorine dioxide and pH. One analog output may be assigned to the pH measurement, allowing remote monitoring and recording of both variables.

 

Flexibility

Programmable range options from 0-200 PPB up to 0-200 PPM provide maximum application flexibility.

 

AC or DC Power Options

Power options include universal 100-240 VAC or 12-24 VDC.

 

Analog Output Options

Two isolated 4-20 mA outputs are standard, with an option for a third output if required. Default setting provides analog outputs for ClO2 and temperature.

 

PID Output

Standard PID control function assignable to one analog output.

 

Digital Communications

Available for either Profibus-DP, Modbus-RTU, or Ethernet-IP.

 

Relay Contacts

Three SPDT relays are standard, with relay functions programmable for alarm, control, or trouble indication. Three additional low power relays available as an option.

 

Flexible Mounting

NEMA 4X (IP-66) enclosure is suitable for wall, pipe, or panel mounting.

 

Clear Display

Back-lit large LCD display provides clear visibility in any lighting conditions. A scrolling second line on the display provides additional information and programming prompts.